Maintenance
SPECTRO iSORT — 23.08.2013
— 25
6.2
Cleaning the adapter
During the measurements, the adapter gets soiled.
After completing work, clean the adapter on the inside and outside, using a cloth
free of lint, oil and grease. Commercially available cotton swabs are also useful for
cleaning.
6.3
Cleaning the light inlet windows of the sample probe and of
the adapter
To clean the light inlet windows proceed as follows:
1. Loosen the adapter (1) using the
knurled screw (2) and pull it off.
Î
Clean the light inlet window
(3) of the sample probe with
a cotton swab.
2. Take the adapter and push the
window holder (1) carefully out
of the guide.
3. Remove the light inlet window(s)
(1) and the exposed O-ring (2).
Î
Make sure not to touch the
optical surfaces.
4. Clean light inlet window using
optical paper or a lint-free soft
cloth.
5. Refit the light inlet window.
Î
Make sure that the O-ring fits in properly under the light inlet window.
6. Re-attach the adapter to the sample probe.
3
The cleaning of the light inlet windows of the sample probe and of the adapter
is now completed.