SERVOPRO CHROMA Gas Analyser
121
04400001A / Revision 7
12.2 FID Advanced Configuration - FID Menu
The FID Menu is used to modify the FID settings. The modifications must be
saved for the modifications to take effect, by clicking on the
Save
button.
Figure 56: Advanced Configuration FID menu
Parameter
Range
Unit
Description
Polariz.
0-200
Volt
Voltage that is applied to the FID polarizer.
Pressure
300-500
In of H
2
O
This value is used to calibrate the pressure sensor
used by the FID. When
Save
is clicked, the
pressure sensor is calibrated with the value entered
in the box.
Carrier*
0-100
cc/min
This carrier set point is only used to ignite the
FID. This set point is not used during normal
operation.
Air*
0-1000
cc/min
This air set point is only used to ignite the FID.
This set point is not used during normal operation.
Fuel*
0-200
cc/min
This fuel set point is only used to ignite the FID.
This set point is not used during normal operation.
Steps
0-10
Steps
Number of steps for ramping flow to normal value
Seconds
0-60
Seconds
Time between each step.
Table 31 : Advanced FID menu parameters