2.4
Sample Gas
The sample gas must have a minimum purity of 5N (<10 ppm).
2.5
Purge Gas
The purge gas must have a minimum purity of 5N (<10 ppm).
Recommended gases to be purified
Ar, He, Ne, Xe, Kr,
Recommended input purity
5N minimum (<10 ppm)
Expected output purity
8N minimum (<10 ppb)
CAUTION
The caps fitted to the gas inlet and outlet must be kept on until just prior to
installation. The length of time that the purifier is left open to the air must be
kept to a minimum, or the getter material may be prematurely depleted.
Recommended purge gas
Gas to be purified in process
Recommended input purity
5N minimum (<10 ppm)
Inlet/outlet flow rate
100 ml/min
Maximum inlet pressure
1000 kPa gauge (145 psi gauge)