![Spectris PARTICLE MEASURING SYSTEMS MiniCapt Скачать руководство пользователя страница 186](http://html.mh-extra.com/html/spectris/particle-measuring-systems-minicapt/particle-measuring-systems-minicapt_operation-manual_1344052186.webp)
Input Registers
Appendix A
: Communications
MiniCapt® Mobile Operations Manual
A-5
Data Section
The
Data
section contains:
•
Calibration Date
•
Serial Number
•
Device Status
•
Device State
•
Number of data samples in queue
•
Sample Data information
The
Data
section starts with what would have normally been considered configuration
information - calibration date and serial number. This was added to allow for a single
contiguous read when this particular information was desired along with the regular
sample data. If not desired, then start reading at register 30212.
30201:
Calibration Date (Month)
A value between 1 and 12.
30202:
Calibration Date (Day)
A value between 1 and 31.
30203:
Calibration Date (Year)
A value representing the year (e.g., 2012).
30204 – 30211:
Serial Number
Serial number entered using ASCII bytes and positioned as shown in the map.
30212 – 30213:
Device Status
Dual-register representation of device status. Integer bit mask matching the
current coil selections for convenience. The host can use this to check for
data availability, data format, etc while reading the data. A separate coil read
would not be required. Values are defined in
30214:
Device State
Least Significant byte (LSB) is State. Most Significant Byte (MSB) is sub-state.
Values are defined in
30215:
Number of data samples in
the queue
A value of zero is identical to no data available (same meaning as coil 00/02
read as zero).
The queue size setting will determine if the data register section
shows real-time data or stored data. If the queue size setting has
been set to one, the sample data will always represent the last
real-time data packet and the queue value will be set to one if
there is a data packet available. If the queue size setting is set
greater than one, the sample data will represent the last data
queued and the queue value will be set to one if there is any
queued data available.
The two types of data presentation are supported by the same
Modbus register map and are delineated by how the host system
processes the register/coil fields and the instrument queue
setting. The data register section can be forced to show real-time
data by setting the queue size to one or to show queued data by
setting the queue size greater than one. Queued data is always
presented oldest first. The queue size command is “set queue #”
where # is a value between 1 and 3000.
Содержание PARTICLE MEASURING SYSTEMS MiniCapt
Страница 13: ...xii MiniCapt Mobile Operations Manual Table of Contents This page is intentionally left blank ...
Страница 81: ...5 10 MiniCapt Mobile Operations Manual Chapter 5 Recipes Delete a Recipe This page is intentionally left blank ...
Страница 197: ...C 2 MiniCapt Mobile Operations Manual Appendix C 有毒或有害的物质和元素 This page is intentionally left blank ...