Configuration Defined
Chapter 3
: Configuring for Ethernet
Airnet® II 4-Channel Particle Sensors Operations Manual
3-5
Select the Modbus TCP mode to interface to any third party Modbus TCP client. For
additional information, refer to Appendix B,
.
Set sam (ple) x y
This command sets the minimum sample interval to
x
seconds and the maximum
sample interval (used when zero particles are detected) to
y
seconds. It is not required
to set the sample interval when running the Airnet II 4-Channel particle sensor with
Facility Net.
Write
Saves the most recent changes.
You must send this command
after setting up the
instrument, or your parameters will be lost when the sensor is de-energized.
TCP/IP Settings for the Airnet II 4-Channel Particle Sensor
:
Viewing Configuration Settings
Type
sta
(
tus
) and press the
Enter
key to display the Airnet II 4-Channel particle
sensor’s configuration settings, software version, and connection status.
If this command does not produce any kind of activity on the display screen:
•
the terminal emulator may be set up incorrectly, or
•
the cable may be improperly connected or faulty.
NOTE:
When the programming is complete, record these settings for
future reference.
Unit Serial Number:
Date
Software Version
Set By
MAC Address
IP Address
Multicast Address
Net Mask Address
Gateway Address
NTP Address