User’sGuide 

 
 

Shop online at 

omega.com 

e-mail:  [email protected] 

For  latest  product manuals: 

www.omega.com 

Above items are included with the DBCL130 
Dry Block temperature calibrator. Not shown 

is the calibration certificate and operator manual 

 
 
 
 
 
 
 
 
 
 
 

Model DBCL130 

Dry  Block  Temperature  Calibrator

 

Содержание Omega DBCL130

Страница 1: ...ail info omega com For latest product manuals www omega com Above items are included with the DBCL130 Dry Block temperature calibrator Not shown is the calibration certificate and operator manual Model DBCL130 Dry Block Temperature Calibrator ...

Страница 2: ...all relevant electromagnetic interference and electrical safety regulations Specification Figures quoted are at the base of the well at the time of calibration Temperature range 25 C 13 F to 130 C 266 F at an ambient of 20 C 68 F NOTE the typical minimum achievable temperature is 45 C 80 F below the room ambient temperature Display resolution 0 1 Accuracy 0 4 C 20 to 130 C 0 7 F 4 to 266 F Stabili...

Страница 3: ... All Omega Engineering units have been designed to conform to international safety requirements If a safety problem is encountered switch off at the power socket and remove the plug from the supply Please use caution when removing probes and inserts as burns to the skin can occur if in contact Installation 1 All Omega Engineering units are supplied with a power cable 2 Before connecting the power ...

Страница 4: ...ue required Alternatively you can press the PF key to move over to individual digits to set higher values much quicker After 2 seconds your value will be set retained 2 When you have entered a set temperature the unit will start to heat or cool to that value 3 Once the process value temperature reaches the set point allow the block to fully stabilize for at least 10 minutes before performing a cal...

Страница 5: ...n certificate CN5 is a calibration offset low end adjustment Example if your reference thermometer reads 0 5 and the DBC130 display is indicating 0 0 then set the CN5 value to 0 5 Then press the first button on the left The display will correct to 0 5 and start to cool down so the block reading and your reference thermometer match Adjust parameter CNRt at high end temperatures to correct non linea...

Страница 6: ...power cable 4168 Unit carrying strap 4285 Insert extractor DBCL 3052 Soft sided carrying case DBCL 130 3074 Insert for 1 8 diameter probes DBCL 130 3075 Insert for 3 16 diameter probes DBCL 130 3076 Insert for diameter probes DBCL 130 3077 Insert for 5 16 diameter probes DBCL 130 3078 Insert for 3 8 diameter probes Spare Parts Part Number Description 4224 Peltier 4174 Temperature controller 4147 P...

Страница 7: ...0 EU The Electromagnetic Compatibility Directive 2011 65 EU The Restriction of Hazardous Substances Directive 6 References to the relevant harmonised standards used or references to the other technical specifications in relation to which conformity is declared Reference Date Title EN 60519 1 2015 Safety in installations for electroheating and electromagnetic processing General requirements EN 6100...

Страница 8: ...378 USA Canada only Engineering Service 1 800 872 9436 USA Canada only Tel 203 359 1660 Fax 203 359 7700 e mail info omega com For Other Locations Visit omega com worldwide The information contained in this document is believed to be correct but OMEGA accepts no liability for any errors it contains and reserves the right to alter specifications without notice ...

Страница 9: ...by OMEGA if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to mishandling improper interfacing operation outside of design limits improper repair or unauthorized modification This WARRANTY is VOID if the unit shows evidence of having been tampered with or ...

Страница 10: ...rs M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH Conductivity Equipment DATA ACQUISITION M U Communications Based Acquisition Systems M U Data Logging Systems M U Wireless Sensors Transmitters Receivers M U Signal Conditioners M U Da...