SERVO
PRO NANOCHROME Gas Analyser
84
04416001A / Revision 7
15.
Repeat the above if there are other peaks to be calibrated.
16.
Disable
Span
Calibration
with the
Disable (F3)
button (one button
performs dual functions Enable or Disable, depending on the active
mode).
Figure 40 : Calibration-Span menu
The value set in the
Calib Gas Value
column must be the impurity
value of the calibration gas before the dilution. The software
automatically calculates the dilution values based on the dilution
factor.
Calibration enabled
will be displayed in the
RUN-REAL TIME
CHROMATOGRAM
menu.
It is recommended to perform
the
calibration in
Automatic
injection mode. This will ensure better gas
equilibrium.