Atlas™ Autotouch 40T Hydraulic Presses
51
Option Screen (Opt) Status Conditions
On the blue coloured background area, the (
Opt
)
screen also shows
the status conditions of the press for the
Pump Temp
. (motor/pump
temperature), and the
Pressure Cal.
(pressure calibration) factor. The
values shown are display parameters only and cannot be changed.
Pump Temp.
- If the motor/pump temperature reaches approximately
55°C, the press will be inactivated for pumping a load. A warning
message “
Pump Too Hot
!” will be displayed on the screen. Only when
the pump cools down sufficiently will the message disappear and the
press can be used again to apply a load. If the press motor/pump
continues to overheat, please contact Specac for advice.
Pressure Cal.
– A typical value shown (circa 1.090) is a factory set
calibration factor for load application readings that is a standard factor
for any Autotouch 40T press.
3) Programming Screen (
Pgm
)
If it is a requirement to operate the Atlas™ Autotouch 40T press just
for a simple set load application under a
maintain load
function for an
unspecified time (an auto “top up” of the load is operating) and a fast
release function, then the Simple Operation Program (
PZ
) is used.
(This program is loaded as standard when the press is initially powered
up).
However, the Atlas™ Autotouch 40T press can be programmed for a
sequence of up to 10 (ten) segments (
S#’s
), to create a particular
sequence of events of set load applications (
Maintain
or
Hold Load
functions), specified hold times (from 0.1 up to 99.9 minutes as a
defined hold time or indefinitely - infinity) and varying load release
rates. The maximum number of programs (
P#’s
) that can be stored in
the Atlas™ Autotouch 40T press is 6 (six) – (
P1
to
P6
). (The (
PZ
)
simple program is always stored and can never be edited or
overwritten).
Содержание Atlas Autotouch 40T
Страница 1: ...Atlas Autotouch 40Ton 40T Hydraulic Presses User Manual 2I 25830 3...
Страница 2: ......
Страница 3: ...Atlas Autotouch 40Ton 40T Hydraulic Presses User Manual 2I 25830 3...
Страница 18: ...User Manual 16 Back View Front Safety Guard Closed 9 13 3 15 8...
Страница 19: ...Atlas Autotouch 40T Hydraulic Presses 17 Detail of Electronic Front Control Panel 4 11 12 1 2...
Страница 81: ...Atlas Autotouch 40T Hydraulic Presses 79 13 EC Declaration Of Conformity...