Atlas™ Auto 40T Hydraulic Press
25
indicator in place of the RUN status symbol. The Increase/Set Load
symbol on the Menu Bar will also be illuminated.
The “Rdy” or “RDY” status means that the “Pump up to” load could be
re-applied at this point by pressing the start button (
11
) again.
However, if a different “Pump up to” load is to be applied, press the
accept (
16
) button to take you into the load setting routine.
When the sample is to be removed from the press, wait until the
Display Screen (
4
) status shows “Rdy” or “RDY” before raising the
safety guard (
5
) and removing the sample.
The Atlas™ Auto 40T press is now ready to accept a further sample.
The Safety Guard
The Atlas™ Auto 40T press incorporates a front safety guard
mechanism (
5
) that is crucial for the safe operation of the press. The
front (and rear) safety guards are made from PETG, a glycol modified
polyethylene terephthalate,
which is a clear, transparent material
offering a high impact resistance to contain any work within the
pressing area in the unlikely event there is an immediate release of
stored energy in the sample. When the guard is closed (down) and the
Rdy/RDY status is indicated, the press is enabled to apply a load to a
sample. When the guard is open (up) the press operation is disabled.
(No power to the motor is allowed). However, the “Pump up to” load
can be adjusted with the guard in the open or closed position.
If the electrical power to the press is disconnected, or there is a cut in
the power when a sample has been compressed and is still under load,
always
ensure that the front safety guard (
5
) remains in the closed
(down) position before re-establishing power to the press. The
indications that this event has happened will be that the power switch
(
1
) on the front panel (
2
) is showing the green (on) color bar, but the
electronic Display Screen (
4
) is blank. When the press is re-powered,
as a safety measure, the press automatically dumps any stored load
and any program sequence will have been aborted.
When the Display
Screen (
4
) indicates “Rdy” or “RDY” the safety guard (
5
) can be
opened, or any previously applied program can be started again by
pressing the start button (
11
).
Содержание Atlas Automatic 40 Ton
Страница 1: ...Atlas Automatic 40 Ton 40T Hydraulic Press User Manual 2I 25470 4...
Страница 2: ......
Страница 3: ...Atlas Automatic 40 Ton 40T Hydraulic Press User Manual 2I 25470 4...
Страница 16: ...User Manual 14 4 Diagrams Of The Atlas Auto 40T Press Front View Front Safety Guard Closed 8 9 4 2 1 5...
Страница 17: ...Atlas Auto 40T Hydraulic Press 15 Back View Front Safety Guard Raised 9 17 3 5 19...
Страница 18: ...User Manual 16 Detail Of Electronic Front Control Panel 4 11 12 13 14 15 16 1 2...
Страница 32: ...User Manual 30...
Страница 65: ...Atlas Auto 40T Hydraulic Press 63 13 EC Declaration of Conformity...