DIRECTIONS FOR POWERED USE
All models
WARNING: Always plug the power cable securely into the wall outlet. Make sure
the wall-mounted outlet will accommodate a heavy duty or hospital-grade plug
and that the outlet is in good working order. The plug of the power cord should fit
tightly into the wall outlet. The plug body, the wall outlet, and the wall plate
should not be cracked or chipped. The plug blades should be securely retained in
the plug body. The ground pin of the plug should be intact and secure.
Do not connect the power cord to an extension cord or to a multiple outlet strip. If
the use of extension cords or multiple outlet strips cannot be avoided, use only
heavy duty or hospital-grade connectors that are approved by the facility. Multiple
outlet strips should be mounted on a fixed object to reduce the risk of liquid spills
and physical damage. In addition, if multiple-receptacle outlet boxes are used,
they also should be protected from the risk of liquid spills and physical damage.
All extension cords and multiple outlet strips should be tagged and inspected
routinely.
Do not cover the power cord with a rug or carpet. Rugs or carpets can prevent
normal air flow, which can lead to greater heat built-up. Place the cord in a low or
no traffic area. Check to be sure the motion of the bed does not interfere with the
bed’s power cord or plug.
CONTROL UNIT FUNCTIONS:
ON/OFF:
Ensure On/Off switch is “Off”. Plug power cord into wall outlet.
On/Off indicator light will illuminate in amber, indicating that the unit is
drawing current but not yet powered up.
Press On/Off switch to “ON”. Indicator light will change to green, along with
additional lights on control panel, indicating that the unit is powered up. Unit
will resume the settings it was in when last powered down.
COMFORT LEVEL:
Initially set the “Comfort Level” on the control panel to
softest selection. Adjust for user comfort as desired, using “Softer” and “Firmer”
buttons.