Span-America PressureGuard APM2 Скачать руководство пользователя страница 1

    

P10331 

 

OWNER’S MANUAL

 

 

PressureGuard

®

 APM2

 

 

and 

PressureGuard

®

 APM2 

Safety Supreme

 

 

 

 

 

 

 

Span-America Medical Systems, Inc. 

Greenville, SC 29615 

800-888-6752 

 

 
 

 

Содержание PressureGuard APM2

Страница 1: ...P10331 OWNER S MANUAL PressureGuard APM2 and PressureGuard APM2 Safety Supreme Span America Medical Systems Inc Greenville SC 29615 800 888 6752 ...

Страница 2: ...P10331 ...

Страница 3: ...POWER LOSS PATIENT TRANSPORT 10 HEAD OF BED ELEVATION 10 TROUBLESHOOTING PATIENT COMPLAINTS 11 GENERAL DIRECTIONS 11 BED LINENS INCONTINENCE PADS BED RAILS CPR STORAGE AND TRANSPORTATION ENVIRONMENTAL CONDITIONS FOR USE SERVICE WARRANTY USE IN WOUND CARE CLEANING ROUTINE INSPECTION OF POWER CORDS AND MATTRESS SPECIFICATIONS 15 TROUBLE SHOOTING GUIDE 17 TECHNICAL DESCRIPTION 19 IN HOME STYLE MODELS...

Страница 4: ...P10331 ...

Страница 5: ...nt instruction clarification The symbol below identifies a WARNING or CAUTION A WARNING identifies situations or actions that may have an effect on patient or user safety Ignoring a warning could cause patient or user injury A CAUTION points out special procedures or precautions that persons must obey to avoid equipment damage The symbol below identifies an ELECTRICAL SHOCK HAZARD WARNING EN 60601...

Страница 6: ... critically ill patients or immobility CONTRAINDICATIONS Not recommended for patients for whom rotation or turning is contraindicated such as but not limited to unstable spinal cord injury unstable skeletal fractures requiring immobilization and or skeletal traction physician orders prohibiting rotation or severe posterior burns requiring skin grafts The PressureGuard APM2 series is not for use by...

Страница 7: ... any opening or hose Keep away from sharp objects 5 Do not use outdoors 6 Possible explosion hazard if used in the immediate proximity of flammable gases risk of explosion 7 Use only original spare parts and consumables 8 Keep power cord away from heated surfaces 9 Plug this product into a correctly grounded outlet only 10 Before cleaning unplug unit from its power source Failure to do so could re...

Страница 8: ...Transfer Zones design provides an additional measure of shearing protection in the form of the silicone coated shear minimizing fabric bands located on the underside of the bi directional stretch cover This Shear Transfer Zones design helps Prevent heels sacrum and scapula from digging into surface Glide user back to original position following HOB elevation Protect against damaging effects of mic...

Страница 9: ...prietary fabric technology that protects surface from stains odor causing bacteria mold and mildew Foam Topper Illustration item The Geo Matt style foam topper is a high density medical grade foam The unique geometric design consists of over 800 individual cells each of which acts individually to redistribute pressure to reduce heat and moisture buildup on the skin and to reduce shear to underlyin...

Страница 10: ... patient entrapment issues 2 Hang the control unit on the foot board at the end of the bed using fold out hangers Figure A Connect airlines to control unit by pressing quick connector into port on the side of unit Triangle symbol should face front Figure B Audible click indicates secure connection Note If desired hanger lock strip included item P10064 Fig C can be used to help hold control unit mo...

Страница 11: ...quid spills and physical damage In addition if multiple receptacle outlet boxes are used they also should be protected from the risk of liquid spills and physical damage All extension cords and multiple outlet strips should be tagged and inspected routinely Do not cover the power cord with a rug or carpet Rugs or carpets can prevent normal air flow which can lead to greater heat built up Place the...

Страница 12: ...h on the side of the mattress to select between two patterns of alternation Lateral Rotation therapy With mattress toggle switch set to Lateral Rotation two air cylinders on one side inflate while the two on the opposite side deflate gently rotating the patient approximately 20 degrees to one side After approximately 5 minutes the inflation pattern reverses and the patient is rotated to the opposi...

Страница 13: ...o previous comfort setting and Alternate mode Power Failure and Low Pressure Alarms Audible Alarm On Off When indicator light is on an audible alarm will sound if either the Low Pressure or Power Failure indicator light is on Press button to silence the alarm Alarm can also be toggled off in advance if audible alarm is not desired for low pressure conditions Power Failure During power failure situ...

Страница 14: ...rs providing an even static air surface to protect the user s skin until power is restored WARNING DO NOT MOVE USER ON MATTRESS ONLY Mattress should not be used alone for user patient transport HEAD OF BED ELEVATION All support surfaces using air as a support medium are designed for distributing pressures over the body in a flat horizontal position Bending the support surface and the body at the m...

Страница 15: ...e CAUTION Be careful not to puncture the mattress with needles or sharp instruments This may result in loss of integrity of the cover or internal air system Regularly inspect the mattress cover for cuts rips cracks or tears Do not use the mattress if the cover is damaged BED RAILS Due to concerns over the possibility of patient entrapment Span America recognizes that the use of rails of any length...

Страница 16: ...31 C decreasing linearly by 50 relative humidity at 40 C Mains Supply Voltage Fluctuation up to 10 of the nominal voltage Overvoltage Category II Pollution Degree 2 WARRANTY All models are unconditionally guaranteed against failure due to manufacturing defects under normal use for 18 months USE IN WOUND CARE Use of PressureGuard APM2 models is only one element of care in the prevention and treatme...

Страница 17: ...ed disinfectants are not effective against C difficile and should not be used to disinfect environmental services For further information relative to this organism and infection control in the healthcare setting please refer to www cdc gov ncidod hip Do not puncture the mattress with needles or sharp instruments This may result in loss of integrity of the mattress air system and will void the warr...

Страница 18: ...hrough They can be damaged by bed movement deterioration from use or aging or human or equipment traffic The cord s insulation should be intact and there should be no evidence of bulging stretching crimping cracking or discoloration especially at the ends there the cord is attached to the plug body and the control unit 5 Regularly inspect as parts of the bed frame motor mattress and controller and...

Страница 19: ...Supreme with standard cover 80 L SAF5884 29 Safety Supreme with standard cover 84 L In Home styles see notes page 19 AP28038 29 80 L X 38 W In Home AP27554 RT 29 75 L X 54 W In Home Therapy Right AP28054 RT 29 80 L X 54 W In Home Therapy Right AP28060 RT 29 80 L X 60 W In Home Therapy Right AP27554 CTR 29 75 L X 54 W In Home Therapy Center AP28054 CTR 29 80 L X 54 W In Home Therapy Center AP28060 ...

Страница 20: ...ort lid P10063 Replacement airline only without quick disconnect or transport lid P10065 Replacement hanger set P10066 Replacement internal female plate for air lines P10067 Replacement feet P10061 Replacement electrical cord 16 L hospital grade 3 prong P10064 Hanger locking strip P10070 Elbow Connectors to mattress SERVICE 00 Refresh Kit for In Home non therapy section 4050 Zippered mattress bag ...

Страница 21: ...ng is functioning properly Head of bed is elevated or knees are gatched The degree of patient turn achieved is reduced with elevation of the head of the bed or gatching of the knees Adjust each as necessary to meet patient needs while maximizing turn angle Defective Control Unit Call for Service Patient exceeds weight limit Call Span America for assistance with product selection Mattress not infla...

Страница 22: ...rea Electromagnetic interference caused by the unintentional emission of electromagnetic waves of energy These waves are transmitted through the air at various frequencies which may produce interference such as abnormal functioning to nearby electronic equipment Determine if emissions are causing the interference by turning the equipment off and on If the interference in the affected device subsid...

Страница 23: ...hould be selected with powered air therapy cylinders located in Center of mattress for a user sleeping alone or on Patient Right Side for two users or for a single user who cannot prefers not to move to center of mattress Air cylinders not connected to digital control unit must be maintained at desired pressure through regular adjustment inflation with manual pump kit item SERVICE 00 Powered air t...

Страница 24: ... should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC61000 4 4 2kV for power supply line 1kV for input out line 2kV for power supply line 1kV for input out line Mains power quality should be that of a typical commercial or hospital environment Surge IEC61000 4 5 1 kV line s to line s 2 ...

Страница 25: ...nes may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40 70 MHz b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz a...

Страница 26: ...ion distance according to frequency of transmitter m 150 kHz to 80 MHz P d 2 1 80 MHz to 800 MHz P d 2 1 800 MHz to 2 5 GHz P d 3 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the ...

Страница 27: ...23 PressureGuard APM2Series Preventive Maintenance and Repair Log Date Air Filter Power Cord Mattress Repair Manufacturer Span America Date Purchased Serial C Cleaned OK Okay R Repaired Replaced ...

Страница 28: ......

Страница 29: ......

Страница 30: ...THIS PRODUCT IS PROTECTED BY ONE OR MORE OF THE FOLLOWING U S PATENTS 6 223 369 6 848 138 8 438 682 SPAN AMERICA MEDICAL SYSTEMS INC GREENVILLE SC 800 888 6752 P10331 ...