Troubleshooting
System Troubleshooting
This section describes troubleshooting procedures which can be used in conjunction with, or separately from,
the diagnostics to isolate a failure.
Required Tools/Test Equipment
#1 Phillips screwdriver
Display
Garbled Display, No Display, or Touchscreen Response
1 Ensure that the external DC power supply is properly connected (power LED to the right of the power ON/
OFF switch is illuminated). If the light is OFF, troubleshoot the external power supply.
2 Cycle the power ON/OFF switch several times and verify that the graphic on the power switch illuminates.
If no light is evident, the +5 V power supply on the CPU is not present and will require troubleshooting.
01030C02
KBD did not respond to command
Replace CPU PCBA.
01030C03
KBD failed selftest
Replace CPU PCBA.
01030C04
KBD failed interface test
Replace CPU PCBA.
01030D00
Cannot access AUDIO device registers
Replace CPU PCBA.
01030D01
Could not open audio device
Reboot and retest. If problem
persists, replace CPU PCBA.
01030D02
Audio device did not respond correctly
Replace CPU PCBA.
01030D03
Audio Loopback test failed
Replace CPU PCBA.
01030E00
NVRAM test failed
Replace CPU PCBA.
01030E01
NVRAM clock not running
Replace CPU PCBA.
01030E02
NVRAM checksum error
Zero the NVRAM and reboot.
Re-program all items in the
Biomed
,
CSR
, and
Clinical
menus (Spacelabs field service
engineers only). If problem
persists, replace CPU PCBA.
01030E03
NVRAM read/write memory test failed
Replace CPU PCBA.
Table 1: Diagnostic Failure Messages (continued)
Error Code
Diagnostics Failure Message
Suggested Action