www.spacelabshealthcare.com
1 - 1 2
U
L T R A V I E W
D
I G I T A L
T
E L E M E T R Y
-
S
E R V I C E
M
A N U A L
I
N T R O D U C T I O N
Additional Data Transmitted
The following information is also transmitted:
•
Patient record
•
Low battery indicator
•
Pacer flag
•
Transmitter type and channel number codes
•
Electrode connection status
Transmitter Batteries
The patient-worn transmitters use a 9V lithium or alkaline battery. A
yellow LED flashes when battery level is low.
1
Operational service life (in hours) assuming a new alkaline battery
(minimum 580 mAH capacity) or lithium battery (minimum 1200
mAH capacity) used until the local low battery indicator begins to
flash.
2
NIBP operations from a 90217 ABP monitor sending readings to
the 90343/91343 multiparameter telemetry transmitter. The
90217 ABP monitor will inflate a standard size adult cuff at least
240 times with alkaline batteries.
Table 1-6 Transmitter Battery Service Life
1
(hours)
Battery
Type
9 Volt Alkaline (ANSI/NEDA 1604A)
9 Volt Lithium (ANSI/NEDA 1604LC)
Load
Conditions
2
ECG
Only
ECG and
Con-
tinuous
SpO
2
ECG and 2-
minute
Episodic
SpO
2
ECG and 5-
minute
Episodic
SpO
2
ECG and
30-minute
Episodic
SpO
2
and
NIBP
ECG
Only
ECG and
Con-
tinuous
SpO
2
ECG and 2-
minute
Episodic
SpO
2
ECG and
5-minute
Episodic
SpO
2
ECG and
30-minute
Episodic
SpO
2
90343,
91343
48
24
36
38
40
120
60
100
104
106
90341,
90347,
91341,
91347
52
N/A
N/A
N/A
N/A
132
N/A
N/A
N/A
N/A