Noninvasive Blood Pressure (NIBP) Monitoring
Note:
Use only cuffs specified by Spacelabs Healthcare. Other cuffs may
adversely affect performance and measurement accuracy. Refer to the
élance
Vital Signs Monitor Supplies and Accessories data sheet (P/N
061-2193-00).
Blood pressure measurements can be affected by the position of the
patient, the patient's physiological condition and other factors.
There are no hazards associated with using noninvasive blood
pressure equipment during defibrillation or high-frequency
electrosurgery, because both the cuff and cuff tubing are made of
non-conductive materials.
Oscillometric pressure measurements determined with this device are
equivalent to those obtained by a trained observer using the
cuff/stethoscope auscultation method, within the limits prescribed by
the American National Standard, Electronic or automated
sphygmomanometers.
Th
or
inc
dif
14.1 Phy
Wh
essure (NIBP) monitoring, apply the cuff to the
arm.
S
TEPS
1. Select the proper size cuff. As a general rule, cuff width
the right.
2.
hose to the connector on left side of the
3.
is chapter discusses the NIBP functionality of the
élance
Vital Signs Monit
luding detailed information on setting up the NIBP connection, describing the
ferent NIBP measurement modes, and the different NIBP menu options.
sical Setup Connections
en setting up non-invasive blood pr
:
should span approximately two-thirds of the distance
etween the patient’s elbow and shoulder as shown in
b
the figure on
Connect the
monitor.
Firmly connect the cuff to the hose.
élance
Vital Signs Monitor Operations Manual
Page 14-2