Invasive Blood Pressure (IBP) Monitoring
élance
Vital Signs Monitor Operations Manual
Page 13-4
When complete, the P1 zeroing message disappears.
Note:
If an “Unable to zero” message appears, follow the transducer
lem.
manufacturer's instructions or
your hospital's protocol for specific instructions on removing air from the
system.
Invasive blood pressure systems specified by Spacelabs healthcare are
compatible with high-frequency electrosurgical and defibrillation equipment.
No special precautions are required.
3.
Touch
the
OK key
to confirm.
13.5 Set IBP Label
nce
Vital Signs Monitor if the IBP is set to ART or
.
Touch
the
IBP Numerical Area
.
A Setup IBP window will appear.
manufacturer’s directions to correct the prob
Refer to the catheter, tubing or transducer
The IBP label identifies on the
éla
CVP.
S
TEPS
:
1
2.
Touch
the
IBP Label
key
.
A dropdown list, as shown below, will appear.
3.
Touch
the
required Label
Selection.
The selected label will determine what are the available parameters displayed in
the IBP Setup Window as detailed in the table below.
Label
Physiological Parameter
Displays and Alarms
ART
Arterial Pressure
Systolic, Diastolic, Mean
CVP
Central Venous Pressure
Mean
RAP
Right Atrial Pressure
Mean
LAP
Left Atrial Pressure
Mean
PRS
Generic Pressure
Systolic, Diastolic, Mean