SpO
2
Monitoring
élance
Vital Signs Monitor Operations Manual
Page 12-11
Note
:
Advanced SpO
2
Setting
function is accessible only for Masimo and
Nellcor
SpO
2
.
3.
Touch
the
Sensitivity Mode
key.
A dropdown list, as shown below,
will appear
4.
Touch
the
required Mode
.
.
Touch
OK
to confirm or
Cancel
to quit.
12.8.2.2
FastSat Settings and Data Averaging Settings
FastSat setting enables rapid tracking of arterial oxygen saturation changes. It
is recommended in any clinical environment that requires the most rapid
response for either treatment or assessment. It is not recommended for routine
use as the increased fidelity will also increase the frequency of alarms.
Data averaging feature smoothes the oximetry saturation value by averaging
patient input values over several seconds. The averaging time can be
manually selected from 2 to 16 seconds.
FastSat is automatically enabled whenever data averaging is set to 2-4 or 4-6
o
S
TEP
1.
Touch
the
SpO
2
Numerical Area
.
A Setup SpO
2
window will appear.
2.
Touch
the
Advanced SpO
2
Setting
key
for FastSat and Data Averaging
setup.
3.
Touch
the
Data Average Time and Fast Sat
5
sec n s.
d
S
:
key
.
A dropdown list,
as shown below, will appear.