Manual Mode
VirTis Wizard 2.0
6
Revision 001, 03/09
© SP Industries, Inc. 2009
Manual Mode Operation
Before operating your system in Manual Mode, verify that the manual control setpoints
are correct in the Manual Mode Settings menu. You can modify these setpoints at any
time during the freeze-dry process. Changes are applied when you return to the
Synoptic screen.
To begin manual operation of your freeze dryer:
1.
Insert product probes, if desired, and load the prepared product into the
chamber.
2.
Verify that the chamber door(s) and any valves are closed.
3.
Press
Vacuum
and allow vacuum pressure to reach approximately 500 Torr,
500 mbar or 50 kPa. Press
Vacuum
again to disable the vacuum pump.
Verify that the system is tightly sealed.
4.
Press
Freeze
and allow the product to reach the desired temperature. If you
are not using product probes, allow the shelf temperature to reach and hold at
the Manual Freeze setpoint until product is uniformly frozen.
5.
Press
Condenser
and allow the temperature to reach approximately -40 °C
or 20 °C below the shelf/product average temperature, whichever is colder.
6.
Press
Vacuum
and allow vacuum pressure to reach an acceptable value
(generally less than 500 mTorr, 500
μ
bar or 50 Pa).
7.
Press
Shelf Ctrl
to begin heating to the Manual Heat setpoint. Adjust the
setpoint as desired to optimize primary drying.
Note:
Use the
Shelf Ctrl
button together with the
Condenser
button to control the
shelf temperature at the setpoint.
8.
If vacuum control is desired, press the
Vacuum Ctrl
button. Adjust the
vacuum control setpoint as needed to optimize primary drying.
9.
To perform manual secondary drying, adjust the shelf/product temperature
and vacuum control level as desired.
10.
When the process is complete, turn off all Manual Mode functions in reverse
order.
11.
Press
Vac Release
to break vacuum. If stoppering your product, you may
press
Vac Release
again while the system is equalizing to close the valve at
the desired pressure. After stoppering your product, select
Vac Release
and
allow the system to reach atmospheric pressure.
Note:
There is no manual backfill setpoint. You must use the
Vac Release
button.
12.
Remove the product.
13.
Press
Defrost
to defrost the condenser. When defrosting is complete, remove
all condensate. Rinse and dry chamber and condenser surfaces.