Sentry 2.0 Controls
VirTis BenchTop 2K, 4K and 6K
18
Revision 004, 12/09
Interstage Control
If your unit is equipped with a cascade refrigeration system, you can control the
operation of the second compressor from the Interstage Control screen. A cascade
system is a safety feature used to protect the refrigeration system from overloading.
∧
Interstage Control
Interstage ON
ON –12 DIFF 025
Current Temp 1.3
°
C
If the Interstage Control is set to ON, the
COND
or
AUTO
buttons will activate the
high-stage compressor. The Interstage Control’s Current Temp is monitored until the
Interstage ON temperature is reached. Once the Interstage ON temperature is
reached, the low-stage compressor is activated.
The DIFF temperature is used to prevent overloading of the high-stage compressor. If
the Current Temp exceeds the sum of the Interstage ON temperature and the DIFF
value, the low-stage compressor is deactivated. It will reactivate when the Current
Temp recovers.
Calibration
The Calibration screen allows modification and control of condenser temperature and
vacuum settings. The following procedures require removal of the electrical box cover
under the unit’s right-side panel.
CAUTION!
REMOVAL OF THE ELECTRICAL BOX EXPOSES HIGH-VOLTAGE COMPONENTS.
USE EXTREME CAUTION AND FOLLOW ELECTRICAL SAFETY PRACTICES.
∧
Calibration
Temperature Span 070
Temperature Zero 14.5C 128
Vacuum 760 Torr 128
Condenser Temperature Calibration
To calibrate the condenser temperature, connect a thermocouple temperature
simulator to the condenser temperature input. Set the simulator temperature to 25 °C.
The current temperature is displayed on the Calibration screen. Adjust the numerical
value of the Temperature Zero as appropriate and press
MENU
to enter the value.
Upon proper adjustment of the Temperature Zero, change the temperature simulator
value to -60 °C. Adjust the Temperature Span value so that the temperature displayed
equals the calibrator set value. Recheck the value at 25 °C and repeat adjustment if
necessary. Check calibration at desired temperature intervals for verification.
© SP Industries, Inc. 2009
Содержание VirTis BenchTop 2K
Страница 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Страница 2: ......
Страница 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Страница 10: ......
Страница 18: ...Installation and Setup VirTis BenchTop 2K 4K and 6K 8 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 52: ...Appendix B Serum Vial Capacities VirTis BenchTop 2K 4K and 6K 42 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 54: ...Appendix C Stainless Steel Cleaners VirTis BenchTop 2K 4K and 6K 44 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 56: ...SP Industries 3538 Main Street Stone Ridge NY 12484 USA www SPIndustries com 800 824 0400 845 687 0071 ...