VirTis BenchTop 2K, 4K and 6K
Sentry 2.0 Controls
Revision 004, 12/09
15
© SP Industries, Inc. 2009
Temperature Units
The Temperature Units field displays the current setting for indicating temperature
values. Temperature can be displayed in either Fahrenheit (°F) or Celsius (°C).
Vacuum Units
The Vacuum Units field displays the current setting for indication of system pressure.
System pressure can be displayed in millitorr (mTorr), microbar (
μ
bar) or Pascal (Pa).
Relay Test
The Relay Test screen can be used to troubleshoot electrical or mechanical problems,
or to observe active outputs during a freeze-drying run.
∧
Solid State Relay Output Test
SSR1 SSR2
-Vacuum -Purge
-Comp2 -Bath
-Comp1 -Spare
-Defrost -Heat
∧
-SSR1
∨
-SSR2 Menu-Exit
To test a relay bank, simply press the
Up or Down (+/-)
button and each of the
outputs of the selected relay bank will fire for approximately two seconds each.
Relay positions preceded by a highlighted box indicate the output is active.
Vacuum Pump Maintenance
The Vacuum Pump Maintenance screen is used to adjust vacuum maintenance alarm
parameters and maintenance time durations.
∧
Vacuum Pump Maintenance
Alarm Setpoint: 240
Current Hours: 0000 min: 007
The Alarm Setpoint is the number of hours of continuous operation the vacuum pump
can run before the vacuum pump maintenance alarm will be generated. This can be
used to draw attention to necessary vacuum pump maintenance operations, such as
oil changes. The default factory value is 240 hours, but can be changed by pressing
and holding the
MENU
button while Alarm Setpoint is highlighted. Once prompted,
use the
Up/Down (+/-)
buttons to change this value.
The Current Hours field can be reset at any time by pressing and holding the
MENU
button while Current Hours is highlighted.
Содержание VirTis BenchTop 2K
Страница 1: ...SP INDUSTRIES INC Freeze Dryer Operator s Manual VirTis BenchTop 2K 4K and 6K Freeze Dryers ...
Страница 2: ......
Страница 6: ...iv Revision 004 12 09 SP Industries Inc 2009 ...
Страница 10: ......
Страница 18: ...Installation and Setup VirTis BenchTop 2K 4K and 6K 8 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 30: ...Sentry 2 0 Controls VirTis BenchTop 2K 4K and 6K 20 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 36: ...Basic Operation VirTis BenchTop 2K 4K and 6K 26 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 44: ...Optional Components VirTis BenchTop 2K 4K and 6K 34 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 52: ...Appendix B Serum Vial Capacities VirTis BenchTop 2K 4K and 6K 42 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 54: ...Appendix C Stainless Steel Cleaners VirTis BenchTop 2K 4K and 6K 44 Revision 004 12 09 SP Industries Inc 2009 ...
Страница 56: ...SP Industries 3538 Main Street Stone Ridge NY 12484 USA www SPIndustries com 800 824 0400 845 687 0071 ...