FTS Systems LyoStar™ II
General Maintenance
Rev 012, 04/11
55
© SP Scientific 2011
Calibration of Product Thermocouples
From the active Maintenance screen you can access the calibration screen.
Each thermocouple and the shelf inlet and outlet RTDs should be placed in a NIST
traceable ice bath or each thermocouple jack should be provided with a millivolt
source at the thermocouple plug. Click on the Auto-calibration box.
To calibrate the machine, place in ice bath/millivolt source and press the Auto-
calibration button. The system will zero itself out. The operator can manually dial in
the machine by clicking on the corresponding box and entering the count value. This
is a dimensionless number from -4095 to 4095. After the values are entered, one can
hit the Manual Calibration button and the system will accept the values. After
calibrating, one may save to the disk to load at a later date.
Содержание LYOSTAR II
Страница 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Страница 2: ......
Страница 6: ...iv Rev 012 04 11 SP Scientific 2011...
Страница 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Страница 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Страница 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Страница 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Страница 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Страница 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Страница 71: ......