FTS Systems LyoStar™ II
Optional Liquid Nitrogen Trap-Auto Filling
Rev 012, 04/11
35
© SP Scientific 2011
Cleaning
Make sure that the trap is off line and that power has been turned off on the
1.
LN2 level controller.
Break vacuum by removing the drain plug.
2.
Carefully raise the sensor
3.
Note: The sensor tube material is fragile. Use care when raising and lowering the level
sensor.
Loosen the black thumb screws and raise the lid up the shaft. Do not lift the
4.
lid off the top of the shaft.
Retighten the thumb screws to hold the lid in the elevated position.
5.
The inner reservoir can removed and cleaned.
6.
Drain the condensate using the drain line then wipe out the outer reservoir.
7.
Once cleaned, loosen the thumbs screws again and lower the lid.
8.
Align and secure the latches.
9.
Carefully lower sensor all the way back into the reservoir.
10.
Содержание LYOSTAR II
Страница 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Страница 2: ......
Страница 6: ...iv Rev 012 04 11 SP Scientific 2011...
Страница 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Страница 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Страница 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Страница 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Страница 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Страница 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Страница 71: ......