FTS Systems LyoStar™ II
Optional Liquid Nitrogen Trap-Auto Filling
Rev 012, 04/11
33
© SP Scientific 2011
Operation
Read and completely familiarize all personnel with Safe Handling and Storage
1.
of Liquified Gases before proceeding. Also, review the Liquid Level Controller
Manual.
Verify that all three latches have been closed and that the reservoir cover is
2.
secure and that the level sensor is in position. Sensor should be inserted as
far as it will go.
Note: If sensor is not fully inserted, liquid nitrogen will overflow out of the vent.
Note: The vent line should not be obstructed in any way to avoid pressure build up.
Open the supply valve (where applicable).
3.
Turn on the level controller and allow time for the reservoir to fill (The low
4.
level alarm will need to be muted until liquid level rises above the end of the
sensor).
Note: If cover is not seated, the proximity sensor will prevent the solenoid from
opening.
When the liquid level reaches 4 inches, the controller will shut the solenoid.
5.
To put the trap on-line, open the inlet and outlet valves and close the bypass
6.
valve.
Содержание LYOSTAR II
Страница 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Страница 2: ......
Страница 6: ...iv Rev 012 04 11 SP Scientific 2011...
Страница 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Страница 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Страница 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Страница 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Страница 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Страница 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Страница 71: ......