Quick-Check-PowerSaver Manifold 6 Port South-Tek Systems
Revision: 0
Revision Date: 8/2/17
Page 21 of 30
Green = Zone within tolerable purity limits
6.2
ZONE FLOW STATUS
The flow status of each zone is indicated in the “Zones with No Flow:”
row. Zones with too little or no flow from the APS will show up as a red
letter, whereas zones with sufficient flow will be green.
6.3
CURRENT MODE
The current mode will display one of two things:
Purging – Purging occurs during initial start-up and in-between sampling of zones.
Zone sampling – Displays which zone is being currently sampled
6.4
ALARM STATUS
The alarm status is determined by alarms being triggered. If an alarm has been triggered, the alarm status will read
“Active”. The type of alarm can be viewed by accessing the “Alarm Info” screen in the “Menu”.
6.5
ZONE SETUP
The Quick-Check starts sampling the zones one by one at the same time every day starting with “Zone A”. To change
this time, access “Zone Setup” within the “Menu” screen and press “Sampling Settings”. On the left side of the screen
under the “Time of Day for Sample”, input the desired time for the zones to be read each day it is set to sample.
(Factory default is set to sample each zone daily for “x” minutes, and to purge any zones which are out of spec for 24
hours.)
The duration of purge after the sample and sample frequency may also be changed on this screen. (Factory default is
set to sample every day and to purge for 24 hours)
Figure 15: Home Screen
Figure 17: Zone Setup Screen
Figure 16: Sample Settings