36
high voltage
LOWR:
OFF~0.1µA~10.00mA Current low limit value of DC
withstanding voltage
TIME:
OFF~0.1~999.9S
Test time of DC withstanding
voltage
ARC:
OFF~0.1~10.0 mA
Current Max. value of DC arc
RISE:
OFF~0.1~999.9S
Voltage rising time of DC high
voltage
FALL:
OFF~0.1~999.9S
Voltage falling time of DC high
voltage test
WAIT:
OFF~0.1~999.9S
DC charge waiting time
CHEK:
ON OFF
DC capacitance load charge
current test.
4.3.3
Insulation resistance test parameter setup:
STEP:
01/01
IR
SCAN:
1X2X3X4X5X6X7X8X
VOLT:
0.050
kV
RISE:
0.5
S
LOC
LOWR: 1.0
M
Ω
FALL:
0.5
S
OFST
UPPER: OFF
SAGC:
OFF
RMT
TIME:
0.5
S
ERR
RANG:
AUTO
F1
F2
F3
F4
F5
Figure 4.3.3 IR setup interface
Instruction of IR test parameters:
VOLT:
0.050~1.500kV
Voltage value of IR test
UPPER:
0.1M~50.00G
High limit value of IR
LOWR:
OFF~0.1M~50.0G
Low limit value of IR.
TIME:
OFF~0.1~999.9S
Test time of IR.
RANG:
AUTO
Auto range mode. The range will be
changed according to the test value.
300nA,3µA,30µA,
300µA, 3mA, 10mA
Fixed range mode. Use I = U/R to
estimate the range to accelerate the
test speed and improve the test
accuracy.
RISE:
OFF~0.1~999.9S
Rising time of insulation voltage.
FALL:
OFF~0.1~999.9S
Falling time of insulation voltage.
SAGC:
ON, OFF
Software auto voltage control.