Chapter 5 Measurement procedure and examples
This chapter covers basic measurement procedures and examples. It also offers various
measurement hints.
5.1Clear (CORRECTION OPERATION)
To prevent the stray capacitance from disturbing the measurement accuracy, you must do the
OPEN/SHORT CORRECTION. You can choose one of the methods below.
5.2 Frequency sweep clear
a)
Press [LCRZ], and press the softkey MORE1/3. MEAS SETUP, CORRECTION,
LIMIT TABLE, LIST SETUP, MORE2/3 will be displayed in the softkey area.
b)
Press CORRECTION. The CORRECTION page will be displayed.
c)
Move the cursor to OPEN field. ON, OFF, MEAS OPEN will be displayed in the
softkey area.
d)
Keep the test fixture open, press MEAS OPEN, wait until the bottom line prompts
―OPEN measurement completed‖.
e)
Press ON, enable the open correction function.
f)
Connect shorting bar (ST26010) to the test fixture.
g)
Move the cursor to SHORT field. ON, OFF, MEAS SHORT will be displayed in the
softkey area.
h)
Press SHORT CORRECTION, wait until the bottom line prompts ―SHORT
measurement completed‖.
i)
Press ON, enable the short correction function.
j)
Move the cursor to LOAD field. ON and OFF will be displayed in the softkey area.
k)
Press OFF. Close the load correction function.
l)
Move the cursor to FREQ 1 field. ON, OFF, MEAS OPEN, MEAS SHORT and
MEAS LOAD will be displayed in the softkey area.
m)
Press OFF, close the dot frequency clear function for frequency 1.
n)
Move the cursor to FREQ 2 field. ON, OFF, MEAS OPEN, MEAS SHORT and
MEAS LOAD will be displayed in the softkey area.