ST2515 Operation Manual
12
3.2
<COMP DISP>
Touch <MEAS DISP> or press [MEAS] to select the COMP DISP soft key to enter into the <COMP
DISP> page shown as figure 3-2.
Figure 3-2
The following parameters can be set on this page:
FILE
TOOLS
COMP ON/OFF
COMP MODE
NOM (nominal value)
Low limit (%)
3.2.1
File Manage
1)
Touch FILE, the following soft keys will be displayed:
File manage: Touch this soft key to enter into the file manage page.
Screen copy: Touch this key to save the current screen page to a U disk. If no U disk is found,
this function will be disabled.
2)
Refer to 4.2 for more details about FILE MANAGE.
3.2.2
Tools
Touch <TOOLS>, the following soft keys will be displayed.
DISP (ON/OFF): ON is to display the measurement result and OFF is not to display.
BEEP (NG/GD/OFF): NG: The beeper will sound when the measurement result is not
consistent with the setting value. GD: The beep will sound when the measurement result
conforms to the setting value. OFF: Turn off the beeper.
COUNT (ON/OFF): ON: Start count; OFF: Stop count.
COUNT CLEAR: Clear the count numbers of TOT, IN, HI and LO and start count from zero.