ST1778 DC Bias Current Source adopts high-performance MPU and can provide a constant current
ranging from 0A to 20A. The maximum current output can reach 0-120A by connecting up to five slaves.
ST1778 can be used with most of Sourcetronic
’s L meters and LCR meters. ST1778 provides a
convenient and practical DC Bias Current Source for AC+DC overlap inductor measurement and
magnetic material analysis.
1.2 Basic Accuracy
1.2.1 Accuracy Test Requirements
Within two year since factory calibration;
Environment temperature: 25
5°C;
Relative Humidity: ≤80%;
Power supply is regulated by a line-voltage regulator;
Warm up time: more than 15 mins.
1.2.2 Current Accuracy
Range (A)
Resolution (mA)
Accuracy
0.000
~
1.000
5
±(1%+5mA)
1.025
~
5.000
25
±2%
5.1A
~
20.0*n (n=1~6)
100
±3%
1.3 Accessories
Current output clip
The standard test clip and output line is ET-07-2 and ET-54. (Note: when the current is greater than 40A,
please do not use it as the current output.)
Footswitch
The standard footswitch is ST1801-001.
Link line
Each ST1778S slave is equipped with a SlaverLink.
Test Fixture
The standard high frequency test fixture is ST26004E-1.