TC6
®
Installation and Operation
2-7
2-9. Emergency Sample
Recovery
If the main power shuts off because of a power failure or system
malfunction, the chamber door will not open. A mechanical over-
ride is provided to allow sample recovery in the case of an emer-
gency.
The mechanical override is located below the air inlet vent in the
front panel, as shown in figure 2-3. To operate the override, insert
the end of the override key (PN 78170 supplied), at a slight upward
angle 2/3 of the way into the hole. Carefully move the key left-to-
right until you locate the solenoid shaft. Slowly move the end of the
key downward to contact the washer located at the bottom of the
shaft. With the key pressed tight against the solenoid shaft, use the
key as a lever to apply downward pressure on the washer to release
the chamber door.
W A R N I N G
This procedure is included
for
emergency sample recovery only
and should never be used for any
purpose other than those explained
in this section.
When the main power shuts off, the
brake will not operate. Unplug the
centrifuge plower cord and wait until
the motor stops spinning
before us-
ing the mechanical override.
!
Figure 2-2. Location of the Mechanical Override
2-7/2-8
Содержание TC6
Страница 35: ...TC6 Printed Circuit Boards Schematic Diagrams 4 5 4 6 Figure 4 2 System Wiring Diagram SN 9501733 and below ...
Страница 101: ...Illustrated Parts SORVALL Centrifuges 8 2 Figure 8 1 TC6 Main Assembly ...
Страница 103: ...Illustrated Parts SORVALL Centrifuges 8 4 Figure 8 2 TC6 Chassis Assembly ...
Страница 105: ...NOTES ...
Страница 106: ...NOTES ...
Страница 107: ...NOTES ...
Страница 108: ...NOTES ...