Reprocessing
36
Ultrasonic Generator | DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US
Cleaning the ultrasonic generator and accessories
7.1 Cleaning the ultrasonic generator and accessories
Requirements
The instrument cable is disconnected from the ultrasonic generator.
The irrigation tube is disconnected from the ultrasonic generator and has been
disposed of.
The foot switch is disconnected.
The power cord and the potential equalization cable are both disconnected from
the main power supply.
Excessive or improper cleaning agents
Excessive or improper cleaning agents may cause damage to the ultrasonic genera-
tor.
Clean the ultrasonic generator in a way that prevents the penetration of liquid or
moisture.
Do not use any explosive or flammable cleaning agents.
Procedure
1. Wipe the ultrasonic generator clean with a moist cloth.
2. Wipe the power cord and the foot switch with a moist cloth.
– There is no longer any visible dirt on the ultrasonic generator or the accesso-
ries.
– The ultrasonic generator and the accessories have been cleaned and are ready
for disinfection.
7.2 Disinfecting the ultrasonic generator and accessories
Requirements
The ultrasonic generator and the accessories have been cleaned.
Procedure
1. Give the ultrasonic generator a disinfective wipe down.
2. Give the power cord and the foot switch a disinfective wipe down.
– The ultrasonic generator and the accessories have been disinfected.