Installation
DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US | Ultrasonic Generator
19
Connecting the ultrasonic generator to the main power supply
5 Installation
The ultrasonic generator may only be put into operation by a qualified person who
has been trained in the proper handling of it.
Improper handling of the ultrasonic generator may lead to life-threatening situations
or substantial material damage. The qualified specialists who are the intended users
of the ultrasonic generator must be instructed in its proper handling.
The ultrasonic generator must be installed on a suitable surface which is adequately
load-bearing and accessible. Carts from Söring are recommended.
5.1 Connecting the ultrasonic generator to the main power supply
A power cord is required to connect the ultrasonic generator to the main power sup-
ply.
If multiple medical electrical devices are being used in the patient's surroundings a
potential equalization cable must be used and IEC 60601-1, chapter 16 followed. The
potential equalization cable eliminates any potential differences between medical
electrical devices in the patient's surroundings.
Main power supply with no grounding conductor
An electric shock resulting from the lack of a grounding conductor in the main power
supply may cause serious injury or death.
Only connect the ultrasonic generator to a main power supply with grounding
conductor.
Damaged power cord
Damaged insulation on the power cord may result in electric shock and cause serious
injury or death.
Check the power cord for damage.
Replace the power cord if damaged.
Use only an undamaged power cord.