sorin Paradym VR 8250 Скачать руководство пользователя страница 1

Implant manual

8250 model
Implantable cardioverter defibrillator

Содержание Paradym VR 8250

Страница 1: ...Implant manual 8250 model Implantable cardioverter defibrillator...

Страница 2: ...diagnostic functions high energy shocks audible alarms 2 INDICATIONS Pacing and defibrillation indications are provided by the American College of Cardiology and the American Heart Association ACC AH...

Страница 3: ...lead The defibrillating paddles should preferably be placed in an anteroposterior position Avoid any direct contact between the defibrillation paddles and the conductive parts of the implanted leads o...

Страница 4: ...NT Pacing and sensing thresholds should be measured at implant ENGLISH 11 Pacing thresholds Acute thresholds should be lower than 1 V or 2 mA for a 0 35 ms pulse width Sensing thresholds For proper ve...

Страница 5: ...coils The result is stored in the device memory 7 2 VENTRICULAR TACHYARRHYTHMIA MANAGEMENT Fast VT treatment Applies detection criteria on fast ventricular tachycardia that are different from those of...

Страница 6: ...Indicator ERI point is reached Confirmed malfunction Burial of the patient for environmental reasons the local regulation may require the explantation of the devices containing a battery supply Crema...

Страница 7: ...below are calculated by taking into account 6 months storage 7 8 years Pacing in VVI mode 100 500 ohm 3 5 V 0 35 ms 60 min 1 one 42 J shock per quarter sensor OFF 7 5 years Pacing in VVI mode 100 500...

Страница 8: ...Values Ventricular post pacing margin mV From 0 to 2 by steps of 0 2 0 8 Response to noise Values Automatic sensitivity on noise Yes No V pacing on noise Yes No 11 2 VENTRICULAR ARRHYTHMIA DETECTION T...

Страница 9: ...5 70 75 80 85 90 95 Ramp decrement per cycle ms 0 4 8 12 16 20 30 40 50 60 Scan decrement per sequence ms 0 4 8 12 16 20 30 40 50 60 Time limit min 0 5 1 1 5 2 2 5 3 3 5 4 Minimum cycle length ms 95 1...

Страница 10: ...under the terms of this limited warranty will be forfeited if the PARADYM device has been opened by anyone other than SORIN These rights will also be forfeited if the device has been damaged by carele...

Страница 11: ...A October 2007 0459 2007 Via Crescentino s n 13040 Saluggia VC Italy Tel 39 161 487323 487095 Fax 39 161 487524 www sorin crm com...

Страница 12: ......