![Sony MG40 Series Скачать руководство пользователя страница 103](http://html.mh-extra.com/html/sony/mg40-series/mg40-series_command-reference-manual_411003103.webp)
(E)
4-15
MG40 Series
Data request
Th
is function is used to re-calculate and output the data.
Since the data request command cannot be used in the latch or pause status, use the memory data output
command.
Request data of all axes
Ethernet
Measurement
mode
The command re-calculates the data, and outputs the data of all the axes.
Command format
R
Return format
Data
Target All
axes
Examples of use
Transmission :
R
Return
:
[00A]02=-123.4567
...
(omitted)
Compatible command
R
Request data of designated axis
Ethernet
CC-Link
Measurement
mode
The command re-calculates the data, and outputs the data of the designated axis.
Command format
r Designated axis
Return format
Data
Target
Designated axis, MG42 hub unit
Examples of use
Transmission :
r[00B]
Return
:
[00B]=3.4567
Compatible command
Designated axis r
For further details, refer to the
.