8
CAUTION
When you dispose of the unit or accessories, you must
obey the laws in the relative area or country and the
regulations in the relative hospital.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts that
conforms to the safety regulations of each country if
applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere). If you have questions on the use of
the above Power Cord / Appliance Connector / Plug,
please consult a qualified service personnel.
Warning on power connection for medical
use
Please use the following power supply cord. With
connectors (plug or female) and cord types other than
those indicated in this table, use the power supply cord that
is approved for use in your area.
*Note: Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked ‘Hospital Only’ or
‘Hospital Grade’.
On cleaning
When the cabinet becomes dirty
Before cleaning the unit, be sure to power the unit off and
disconnect the power plug.
If the body of the unit is dirty, clean it with a soft, dry cloth.
In extreme cases, use a cloth steeped in a little neutral
detergent, then wipe dry. Do not use organic solvents such
as alcohol or thinners, as these may cause discoloration or
other damage to the finish of the unit.
For the customers in Canada
This unit has been certified according to Standard CSA
C22.2 No.601.1.
For the customers in Europe
This product has been manufactured by or on behalf of
Sony Corporation, 1-7-1 Konan Minato-ku Tokyo,
108-0075 Japan. Inquiries related to product compliance
based on European Union legislation shall be addressed to
the authorized representative, Sony Deutschland GmbH,
Hedelfinger Strasse 61, 70327 Stuttgart, Germany. For any
service or guarantee matters, please refer to the addresses
provided in the separate service or guarantee documents.
Recommended separation distances between portable and mobile RF communications equipment and the
HVO-1000MD/3000MT
The HVO-1000MD/3000MT is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the HVO-1000MD/3000MT can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (Transmitters) and the
HVO-1000MD/3000MT as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 1.2
√
P
800 MHz to 2.5 GHz
d
= 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated a maximum output power not listed above, the recommended separation distance
d
in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
United States and Canada
Plug Type
HOSPITAL GRADE*
Cord Type
Min.Type SJT
Min.18 AWG
Minimum Rating for Plug
and Appliance Couplers
10A/125V
Safety Approval
UL Listed and CSA