5
Important EMC notices for use in the medical environments
• The HVO-1000MD/3000MT needs special precautions
regarding EMC and needs to be installed and put into
service according to the EMC information provided in
this instructions for use.
• The portable and mobile RF communications equipment
such as cellular phones can affect the HVO-1000MD/
3000MT.
WARNING
The use of accessories and cables other than those
specified, with the exception of replacement parts sold by
Sony Corporation, may result in increased emissions or
decreased immunity of the HVO-1000MD/3000MT.
WARNING
If the HVO-1000MD/3000MT should be used adjacent to
or stacked with other equipment, it should be observed to
verify normal operation in the configuration in which it
will be used.
List of cables used for EMC tests
Type of cable
Specifications
AC mains cable
2 m, non-shielded
Y/C In port cable
2 m, non-shielded
COMPOSITE IN port cable
10 m, shielded
DVI-D Input port cable
10 m, shielded
HD-SDI INPUT cable
10 m, shielded
PARALLEL REMOTE port (8pin) cable
15 m, non-shielded
SERIAL REMOTE port cable
1.5 m, non-shielded
VIDEO OUT port cable
1.5 m, shielded
S-VIDEO OUT port cable
2 m, non-shielded
DVI-D OUT port cable
3 m, shielded
YPbPr OUT port cable
3 m, shielded
HD SDI OUT #1 port cable
1.5 m, shielded
FS/TRIG port cable
2 m, non-shielded
LAN cable
2 m, shielded
USB cable
0.5 m, shielded
RGB cable
1.5 m, shielded
Sound cable
2 m, non-shielded
Sound cable
2 m, non-shielded
Serial cable
3 m, non-shielded
Guidance and manufacturer’s declaration-electromagnetic emissions
The HVO-1000MD/3000MT is intended for use in the electromagnetic environment specified below.
The customer or the user of the HVO-1000MD/3000MT should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The HVO-1000MD/3000MT uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The HVO-1000MD/3000MT is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies