Chapter 5:
Performance Testing
39
Battery Charging
Video Output
Test Battery
Charging
Operation
1
Remove the system from the docking system and insert a battery into the system.
2
Press the
Power
key to turn the system on. Allow the battery to discharge. The battery
indicator icon on the display, below the Transducer Type indicator, will extinguish from left
to right as the battery discharges.
Note: The Power and Sleep delays in the Setup page should be selected to “Off” to properly
perform this test. The battery may take 1–2 hours to discharge.
3
Reattach the system to the Docking System and attach the AC power cord to the power
connector.
4
Note that the battery indicator indicates that the battery is charging. The sections of the
battery indicator will light sequentially from left to right as the battery charges.
Caution:
Use only the recommended video monitor or printer when verifying the video output at the
video receptacle.
Test Video
Output
1
Attach an external video monitor to the video connector using the video cable.
2
Turn on the system power and verify that the video on the external monitor matches the
video on the system display.
If the video does not appear similar, or there is no display on the external monitor, see
Содержание M-Turbo
Страница 1: ...M Turbo Ultrasound System Service Manual TM...
Страница 6: ...2 Chapter 1 Introduction...
Страница 20: ...16 Chapter 3 Troubleshooting...
Страница 36: ...32 Chapter 4 Replacement Procedures...
Страница 44: ...40 Chapter 5 Performance Testing...
Страница 48: ...44 Appendix A Replacement Parts List Figure A 1 Power Supply P08850 Figure A 2 Speaker Assembly P03872...
Страница 52: ...48 Appendix B Service Event Report Service Event Report Form...
Страница 56: ...52 Index...
Страница 57: ......
Страница 58: ...P08144 01 P08144 01...