S-36 VLF
– USER MANUAL
11
3
Preparation for use
3.1 Operating requirements
It is not permissible to leave the operating unit in the protective bag during operation of the unit.
Make sure that the device is adequately ventilated. Do not cover the ventilation slots. After putting
down th
e device carrying handle can be locked in 30°-steps by using spring bias. Pulling on handle
unfixes the lock. The operating unit may be operated either in the vertical position (with the operator
panel facing upwards) or in the horizontal position (with the operator panel facing towards the front).
The connecting cables are kept in the front pocket. When packing in the cables, in particular the
high-voltage connection cable, make sure that it will not be kinked.
The high-voltage unit must always be set up in the vertical position
and secured during oper-
ation against inadvertent contact. Make sure that the high-voltage device is placed on a clean surface
so that it is protected from dirt and moisture and that no foreign matter may get into the air intake and
outlet openings.
WARNING
Strictly ensure that the clearances between the high-voltage terminal and
grounded or live parts are kept in accordance with EN 50191 standard.
NOTE!
During the test, warning signs with the text:
HIGH VOLTAGE!
Caution! Danger to Life!
must be put up, and the area of testing must be cordoned off with a barrier. Set up the
operating unit at a distance of approx. 3 m from the barrier. It should preferably be set
up on a surface which is at operating height (
0.8…1.0 m).