6.8 Measurement of capacitance
WARNING:
Risk of electric shock. Disconnect the power supply from the tested capacitor and discharge
all capacitors before any starting capacity measurements.
To perform the measurement:
set the rotary switch at
Ω
CAP
,
connect
black
test lead to
COM
terminal and
red
test lead to
V
Ω
CAP Hz% Temp
terminal,
select
MENU
and use
arrows
and
F1 F2 F3 F4
buttons to set
CAP
,
contact the probe tips to the tested capacitor,
read the measurement result on the display,
after completing the measurements, remove test leads from the terminals of the meter.
6.9 Current measurement up to 10 A
NOTE
Do not make current measurements of 20 A DC and AC for longer than 30 seconds. Exceeding
this time may cause damage to the meter and/or test leads.
To perform the current measurement:
set the rotary switch at
10A
,
select
MENU
and use
arrows
and
F1 F2 F3 F4
buttons to set the measurement of:
direct current voltage
DC
(default setting)
,
alternating current voltage
AC
,
connect
black
test lead to
COM
terminal and
red
test lead to
10A
terminal,
remove power supply from the measured circuit and then connect the meter in series to the cir-
cuit at the point where the current is to be measured,
contact the probes to the poles of the tested object;
Direct Current
(DC):
the pin of
black
probe should contact
negative
pole of the circuit, while
the pin of
red
probe should contact
positive
pole of the circuit,
turn on the power supply of the circuit,
read the measurement result on the display,
after completing the measurements, remove test leads from the terminals of the meter.
6.10 Current measurement up to 500 mA
To perform the current measurement:
set the rotary switch at
mA
,
select
MENU
and use
arrows
and
F1 F2 F3 F4
buttons to set the measurement of:
direct current voltage
DC
(default setting),
alternating current voltage
AC
,
connect
black
test lead to
COM
terminal and
red
test lead to
µA/mA
terminal,
remove power supply from the measured circuit and then connect the meter in series to the cir-
cuit at the point where the current is to be measured,