Threshold low and high activity [mg]
This value is used for the detection of times with high, median and low activity. Those values will be
displayed for the user defined areas in the report. The activity values calculated in the pre-set
Timeframe are compared with this value. If the calculated value is higher/lower than the respective
Activity Threshold
the period will be counted as
high
or
low activity.
If the calculated value is
between both thresholds, the period will be counted as
median activity
.
6.2.3.4 Flow Analysis
If the Flow signal gets lost during a recording, e.g. due to shifting of the cannula, the Flow
Source can be changed (e.g. to Sum Effort) via a right-click at the raw data at the point of loss.
This allows for a continuous Flow analysis.
Fig. 6.9: Flow Analysis
“Hide events during wake”
If you activate this checkbox, automatically scored flow events during wake are not displayed.
Manually edited events during wake are still shown. When you edit events manually, automatic events
that are hidden, are overwritten by the manual scoring.
1)
Analysis Window [s]
Timeframe used by the analysis.
ACT
ACT