An internal measurement interval can be set for the USH-9. If selected in menu item
, measurements are performed in the defined interval. However, a measurement is always
completed before a new one is initiated.
11.5.2
The menu language.
11.5.3
The character used as decimal separator in the values of the settings and in serial data strings.
11.5.4
The units and number of decimal digits. These have to be set prior to all other settings as all values
are saved internally in this format. They are set in the parameter menu
.
ATTENTION
If units or decimals are changed, related parameters may need to
be adjusted.
11.5.5
The type of the serial output protocol. The following options are available:
ID
Option
Description
1
Sommer (default)
Sommer protocol; data values are returned with an index starting at 1
2
Standard
Standard protocol; data values are returned without an index
3
MODBUS
Modbus protocol
11.5.6
The main measurement values are always included in the data output string. Additionally, special and
analysis values can be included.