B.2.3 Recorded data show frequent data gaps
Reason
Solution
Measured level is out of range.
Verify that level measurements are within the
range of the USH-9.
USH-9 is not perpendicular to the ground
Verify that the USH-9 is mounted perpendicular
to the ground (see also
).
B.3
Firmware & software
B.3.1
Commander loads wrong setup
If the setup is reloaded from the device the Commander seems to display an old version.
Cause
Solution
The device has been con-
nected to the same PC before
and several different setup
files have been loaded.
Delete the setup files of the device that have been downloaded by
Commander to the folder
C:\User-
s\Public\Documents\Sommer\Setup
. The respective files can be
identified by the serial number in the file name and the file date.
B.3.2 Firmware update via RS-485 is aborted
Reason
Solution
USB to RS-485 converter cable is damaged or
can only operate on 9600 baud.
Replace USB to RS-485 converter cable. The pro-
grammer requires 57600 baud.