Somat eDAQ
lite
50
I2773-4.
4
en
3.2.1
Communications
Connect Timeout Period
Specify the desired timeout period in seconds for initiating communications with the
eDAQ
lite
. The default value of five seconds should work well for dedicated ethernet
communications with an eDAQ
lite
that is not on a network hub. Longer timeouts may
be required for communications with an eDAQ
lite
on a busy or slow network hub or
on a wireless ethernet connection.
NOTE
The communications I/O timeout, used for all PC to eDAQ
lite
communications, equals
the connect timeout period for periods greater than ten seconds. Otherwise, TCE sets
the I/O timeout period to the minimum timeout period of ten seconds.
Socket Buffer Size
Specify the buffer size used for PC to eDAQ
lite
communications. In general, leave the
default value of 61440 bytes for optimum communications data throughput. For
situations where timeouts on communications occur, increasing the timeout period is
recommended over decreasing the socket buffer size.
IP Address or Host Name
Specify the IP address or host name of the desired eDAQ
lite
.
3.2.2
General
Descriptions (not IDs) for data file (plot) labels
Store the TCE channel description fields in the data file for use as labels on plot
windows. When selected, empty or duplicate channel descriptions prompt warnings
from TCE. Modifying this option is not available after test initialization.
NOTE
Although TCE description fields allow 63 characters, data file labels are limited to 31
characters. Because the Rainflow and Peak/Valley Matrix DataModes add prefixes to
the labels, it is recommended to limit the description field to 24 characters when using
these DataModes.
Require user to verify test control selections
Force TCE to issue a verification prompt after every Test Control menu command.
Chain auto range options to stop test run
Require TCE to open the auto range options dialog after each test run stops. For more
information on auto range options, see
“Auto Range Options” on page 57
Chain SIF frame demultiplex to upload/consolidate (SIF Only)
Force TCE to demultiplex all multiplexed SIF file data records after successful upload
or SIC consolidation. This option significantly improves data analysis performance by
products such as InField. Note that TCE does not perform the demultiplexing if there
are any abnormalities in the upload or consolidation.
HBM: public
Содержание eDAQlite
Страница 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Страница 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Страница 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Страница 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Страница 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Страница 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Страница 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Страница 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Страница 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Страница 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Страница 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...
Страница 227: ...I2773 4 4 en 227 Somat eDAQlite HBM public ...