Mini Water Level Meter: Op Instructions

Model 101M

Equipment Check

1. Turn Sensitivity Dial fully clockwise.

2. Depress the Battery Test button to 

test the battery and circuitry.

3. Immerse the probe in tap water. This 

completes the circuit and activates the 
buzzer and light.

Routine Care

1. After the depth to water has been 

recorded, the tape should be carefully 
rewound onto the reel, the probe wiped 
dry and placed into the probe holder.

2. The probe, tape and reel can all be 

cleaned with phosphate free 
(non-abrasive) detergent and water.

Battery Replacement

• battery type - alkaline, 9 volt.

Note:

Before replacing the battery, ensure 
the tape is fully wound onto the 
reel, the probe is in the probe holder, 
and the brake has been applied.

Note:

Before screwing the faceplate back
onto the hub, make sure no wires 
are outside the hub.

Notes:

1. Clockwise rotation of sensitivity dial

turns meter on and increases 
sensitivity.

2. Always set switch to highest 

sensitivity position, then decrease 
if necessary.

1. The battery is housed within the 

faceplate of the reel.

2. Using a flat headed screw driver, remove 

the three screws on the faceplate.

3. Carefully remove the faceplate so as 

to not unwind the tape.  Note that 
t

here is an elastic band keeping everything 

together in the faceplate.

4. While leaving the elastic in place, 

replace the battery.

7. Carefully place the faceplate back 

onto the hub and line up the holes.

8. Once the screws have been tightened 

and the faceplate is secure, test the 
water level meter.

Replacement Parts

The following parts can be provided
should they become lost or damaged.

1. Probe and seal kit

2. Splice kits

3. Lights, switches, etc.

4. Reels

5. Replacement tape with probe

SYMPTOM

CAUSE

REMEDY

No sound when probe
immersed in water.

Dead battery.  

Replace with 9V Alkaline.

Water conductivity is very low.

Increase sensitivity switch setting (turn
clockwise) or call Solinst for assistance.

Disconnected wires on circuit
board.

Check all connections inside hub of reel for 
loose/disconnected wires - solder or reconnect.

Broken wire in tape.

Locate break in tape - splice and seal. 
(Contact Solinst)

Disconnected wire inside probe. Contact Solinst to obtain parts/repair instructions.

Instrument continuously
sounds after being
immersed in water.

Water in probe. Probe may be
dirty which could interfere with
the circuit connection.

Contact Solinst for instructions to remove, 
clean and reseal.

Model 101M

Mini Water Level Meter

Georgetown

Ontario

(905) 873-2255

Made in Canada

Battery Test Button

Probe Holder

Sensitivity
Switch

Buzzer (Sonalert)

Winding Handle

LED Light

Carry

Handle

June 22, 2006