Solinst 101B Скачать руководство пользователя страница 1

Water Level Meter: Op Instructions

Model 101B

Printed in Canada 

June 22, 2015
(#112205)

For further information contact:  Solinst Canada Ltd.

Fax: +1 (905) 873-1992; (800) 516-9081  Tel: +1 (905) 873-2255; (800) 661-2023

35 Todd Road, Georgetown, Ontario  Canada  L7G 4R8

Web Site:  www.solinst.com   E-mail: [email protected]

®Solinst is a registered trademark of Solinst Canada Ltd.

Equipment Check

1.  Submerse the probe in tap water to test 

the probe. This completes the circuit 

and activates the buzzer and light.

Routine Care

1.  After the depth to water has been 

recorded, the tape should be carefully 

rewound onto the reel, the probe 

wiped dry and placed into the probe 

holder.

2.  The probe, tape and reel can be 

wiped clean with phosphate free 

(non-abrasive) detergent and water. 

Do not submerge the reel.

Battery Replacement

•  Battery type - alkaline, 9 volt.
1.  The battery is housed within the 

faceplate of the reel.

2.  Using a Phillips screwdriver, remove 

the three screws on the faceplate.

3.  Carefully remove the faceplate so as 

to not unwind the tape.  

Replacement Parts

The following parts can be provided should 

components become lost or damaged.
1.  Splice kits
2. Electronics
3.  Faceplate and backplate
4 Tapes
5.   Probes
6.  Probe holders, brakes, etc.

4.  Replace the battery.
5.  Carefully place the faceplate back 

onto the hub and line up the holes.  

Troubleshooting

SYMPTOM

CAUSE

REMEDY

No sound when probe 

immersed in water.

Dead battery.

Replace with 9V Alkaline.

Water Conductivity is very low.

Try a different Solinst 101 Water Level Meter.

Disconnected wires on circuit 

board.

Check all connections inside hub of reel for  

loose/disconnected wires - solder or reconnect.

Broken wire in tape.

Locate break in tape - splice and seal, or replace. 

(Contact Solinst)

Damaged probe.

Contact Solinst to obtain parts/repair instructions.

Instrument continuously 

sounds after being 

immersed in water.

Probe may be dirty which 

could interfere with the circuit 

connection.

Contact Solinst for instructions to remove, clean 

and replace the probe.

Probe 

Holder

Note: 

Before screwing the faceplate back 

onto the hub, make sure no wires are 

outside the hub. 

6.  Once the screws have been tightened 

and the faceplate is secure, test the 

water level meter.

Water Level Measurements

1.  The zero measurement point on the 

101B P1 Probe is at the tip of the 

conductor pins.

2.  Feed the tape into the well. If a tape 

guide is used, lay the tape onto the 

groove on the top. Measurements will 

be read at the point of the V-notch on 

the tape guide. Remember to deduct 

2/10 ft or 6.0 cm.

3.  The light and buzzer activate when the 

probe conductor pins enter water. To 

ensure accuracy, lower and raise the 

probe a few times and then record the 

depth measurement from the tape at 

the top of the well.

Carry 

Handle

LED Light

Winding 

Handle

Tape Guide (Optional)

Fit onto well casing

Measurement point 

(Deduct 2/10 ft or 6.0 cm)

Sonalert

101B P1 

Probe 