Water Level Meter Circuit Board Replacement

Model 101/102

Tools and Materials Needed

1.  Mk2 Circuit Board Assembly (#102977) Includes:

• Wiring, Sensitivity Switch and Knob, Light and Red  Lens, and Test Button

2.  Phillips Screwdriver
3.  10 mm (3/8”) Wrench
4.  Wire Cutters (if required)
5.  Small Flat Screwdriver (3 mm (1/10”)
6.  Pliers

Back of Mk2 Water Level Meter Faceplate with New Style Light

Instructions  

For Mk2 Water Level Meters (Push-Release Tape Connection)

Note:  

Instructions for Mk1 Water Level Meters (Molex tape 

connection, soldered connections to battery tray) are on Page 2.

1.  Remove the battery. Unscrew the three screws and remove 

the faceplate.  

2.  Press down on the white terminals of the push-release 

fittings on the circuit board and pull out to remove the tape 

leads. 

3.  Pull the quick-connect fittings off of the terminals on the 

battery tray. 

4.  From the front of the faceplate, use the wrench to unscrew 

the nut holding the test button and push the button back 

out of the faceplate.

Back of Mk2 Water Level Meter Faceplate  

with Older Style Light

Light

5.  If you have an older style light (leads connected to Sonalert 

terminals – see photo above), cut the red and black wires 

connected to the light, and push the light out through the 

front of the faceplate. 

6.  Use the small flat screwdriver to unscrew the small brass 

screw on the side of the sensitivity knob. Remove the 

sensitivity knob. Use the pliers to unscrew the nut holding 

the sensitivity switch and remove the switch from the 

faceplate. Save the knob and screw to be used with the 

new sensitivity switch.

7.  Remove the two screws to disconnect the old circuit board 

assembly from the Sonalert. Remove the old circuit board.

8.  Loosen the Sonalert retaining nut on the front of the 

faceplate, so you can reposition the Sonalert as shown in 

the photo below (negative terminal next to the battery tray). 

9.  If you had an older style light (or the lens needs replacing) 

insert the new red lens through the opening in the faceplate 

and secure using the white nut. 

10. Place the circuit board assembly in position on top of the 

Sonalert, ensuring the sides ve and -ve are lined 

up with the corresponding terminals on the Sonalert and 

the light seated in the lens.

11. Install the two screws into the Sonalert terminals. Do not 

tighten. Wrap the two bare wires from the circuit board 

around the screws, between the screw head and Sonalert 

terminals. Start by bringing the wires underneath the 

screws (see photo below). Tighten the two screws.

(Page 1 of 3)

Leads From 

Light Connected 

To Sonalert

Light

Battery 

Tray

Sonalert

High Quality Groundwater and Surface Water Monitoring Instrumentation