9.4 COMMON ERROR MESSAGES & STATUS CODES
9.4.1 Display Error messages
Error code
Description
Action
Internal Error
1
Hardware fault 1
Recycle power if error still occurs unit must be returned
for repair
2
Hardware fault 2
3
Hardware fault 3
Node Error
1
Network Error 1
The Orbit Module on the network can no longer be found
(at positon1, 2 3 or 4).
The module has been changed and needs to be re-
notified or some other problem has occured.
Cycle power and then reconfigure the network.
2
Network Error 2
3
Network Error 3
4
Network Error 4
Probe Error
1
Probe Error 1
An in line connector type probe has failed or been
swapped by the wrong type. Check the probe type
matches the module type. Correct as required. Cycle
power and re-configure network.
2
Probe Error 2
3
Probe Error 3
4
Probe Error 4
9.4.2 Detachable Probes Error Messages
If a compatible probe is not fitted at startup or is dis-connected during use, a pop-up message is displayed stating ' Internal probe
detached or incompatible'. This message is also displayed if a compatible probe is not connected and an attempt to add a probe is
made.
External modules do not generate this error message, however the Reading Status parameter will indicate the appropriate values.
503110 - Orbit ACS Manual
Issue 9
31 of 63
Содержание AMETEK ORBIT ACS Series
Страница 1: ......
Страница 17: ...9 1 SI100 AND SI200 DISPLAYS 503110 Orbit ACS Manual Issue 9 17 of 63 ...
Страница 23: ...9 3 1 Top Level Menu 503110 Orbit ACS Manual Issue 9 23 of 63 ...
Страница 24: ...9 3 2 Configure Probe s 503110 Orbit ACS Manual Issue 9 24 of 63 ...
Страница 29: ...9 3 5 Peak Preset Limits and Misc Menu 503110 Orbit ACS Manual Issue 9 29 of 63 ...
Страница 30: ...9 3 6 Output Input and Communications Menu 503110 Orbit ACS Manual Issue 9 30 of 63 ...
Страница 47: ...15 5 ASCII CHARACTER SET Example ASCII A is HEX 41 503110 Orbit ACS Manual Issue 9 47 of 63 ...