42
5.8 Tenfold display x10
Shows the weight value with another decimal in 10 times higher resolution.
Switch to the setting function level with the function change key.
Gross
x 10
Aux.
display
Prog-
ram
Setup
mode
F 1
F 2
F 3
F 4
F 5
F 6
Press the F2 x10 key.
Approvable scales
If the scale has been calibrated as verifiable, the tenfold higher resolution appears as long as the x10 key is
held down. After the key is released, the x10 display switches off after 5 seconds.
Non-approvable scales
If the scale is not calibrated for use in legal metrology, the tenfold higher resolution appears constantly in the
display. Pressing the key, switches on the tenfold higher resolution. Pressing another key switches the
tenfold higher resolution off again.
As long as the tenfold higher display is switched on, X10 appears in the symbol bar. When the machine is
turned off, the function is disabled.
In the setting mode, you can set the tenfold higher display to be constantly displayed for one weighbridge.
The function setting mode/scale/scale parameters/control mode/on switches the x10 display constantly on
and deactivates the x10 key.
Prerequisite: Scale is not approvable. The factory setting is "Function off".
Содержание 3035
Страница 1: ...www soehnle professional com Manual Terminal 3035 ...
Страница 2: ...2 ...
Страница 122: ...122 ...
Страница 123: ...123 ...