27
For 6 measuring point numbers hardware addressing (system) new user measuring point numbers (user) can
be assigned variably.
Select the desired system address Sys: xx with the arrow keys, then press F6 Continue to enter the desired
new user address Use: xx, then press F6 Apply.
For the user address Use: 00, the machine address is automatically used.
Multiple assignment of the same measuring point numbers is possible.
2
Allocation table measuring points
Sys: 01Use : 10
Sys: 02Use : 00
Sys: 00Use : 00
Sys: 00Use : 00
Sys: 00Use : 00
Sys: 00Use : 00
System no. 01
User no. 10 (new measuring points no. = 10)
System no. 02
User no. 00 (measuring points no. = 02)
System No. 00
User No. 00
System No. 00
User No. 00
System No. 00
User No. 00
System No. 00
User No. 00
Querying the calibration and adjustment points helps service personnel to check the scale.
3
Adjustment bases
Current raw measured value Query of
the current raw measured value (variable depending on the load on the
scale)
Zero point raw measured
value
Query of
the adjusted raw measured value from the zero point
support point 1 weight value Query of
the weight value from 1 base point
Base point 1 raw measured
value
Query of
the adjusted raw measured value from 1 base point
Support point 2 Weight value Query of
the weight value of the 2 support point
Base point 2 Raw measured
value
Query of
the adjusted raw measured value from the 2 base point
Support point 3 Weight value Query of
the weight value of the 3 support point
Base point 3 Raw measured
value
Query of
the adjusted raw measured value from the 3 base point
Support point 4 Weight value Query of
the weight value of the 4 support point
Base point 4 Raw measured
value
Query of
the adjusted raw measured value from the 4 base point
The internal measuring point is on the central board of the terminal 3035 with the measuring point number
01.
4
internal measuring point
Operating status
Query or selection
from or on
With off, the internal measuring point is deactivated with on, measuring point 01 is
active.
Factory setting: on