SNDWAY SW-6510S Скачать руководство пользователя страница 1

ULTRASONIC  THICKNESS  GAUGE

SW-6510S

mm

4.27

Scan

Mat:

5920

Steel

mm

Max

mm

Min

mm

Avg

ULTRASONIC

THICKNESS  GAUGE

Содержание SW-6510S

Страница 1: ...ULTRASONIC THICKNESS GAUGE SW 6510S mm 4 27 Scan Mat 5920 Steel mm Max mm Min mm Avg ULTRASONIC THICKNESS GAUGE...

Страница 2: ...garbage please handle them in line with related laws and regulations 4 If the company s products beyond the warranty period are faulty they can be handed over to the company to repair the product and...

Страница 3: ...ials It can monitor various pipes and pressure vessels in production equipment monitor their thinning degree after being corroded during use and can also accurately measure various plates and various...

Страница 4: ...passes through the couplant to reach the measured object part of the ultrasonic signal is reflected by the bottom surface of the object and the probe receives the echo reflected by the bottom surface...

Страница 5: ...value the MIN value and Average Value Display 2 4 inch black and white dot matrix screen Sound Velocity 1000 9999 m s Lower limit of pipe measurement 20 3mm steel Battery Built in 3 7V 2000mAh Li ion...

Страница 6: ...ance 1 Housing 2 Buttons 3 Display 4 Transmitter socket 5 Receiver socket 6 Thickness calibration block 7 Nameplate 8 Ultrasonic thickness probe probe for short 7 mm 20 00 Scan Mat 5920 Steel mm Max m...

Страница 7: ...ry power 5 Sound indication alarm sound switch display 6 Information display display thickness measurement value measurement material sound speed maximum minimum average The product is powered by a bu...

Страница 8: ...e the inversely measured sound velocity value to measure the thickness of the workpiece of the same material Note When not in use for a long time fully charge the product first and recharge it every s...

Страница 9: ...ss the power button for more than 10 seconds to force shutdown Apply the couplant evenly to the tested area tightly couple the probe to the surface of the tested material and the screen will display t...

Страница 10: ...ement in real time 7 3 Continuous measurement 7 4 Alarm measurement The user can monitor the material with poor quality through the alarm function Press and hold to turn the alarm on or off The displa...

Страница 11: ...measure 1 Press and hold to enter the calibration mode and the screen displays Please measure the calibration thickness block 2 Referring to Figure 1 and Figure 2 in 7 2 Single Measurement apply the...

Страница 12: ...ess to adjust the thickness value of the instrument and adjust it to the same value as that of the caliper 5 Press to calculate the sound speed of the workpiece material 6 At this time the correct sou...

Страница 13: ...lect the speed of sound according to the material When the current unit is mm the unit of sound speed is m s and when the unit is in the unit of sound speed is in us In the display state of the main i...

Страница 14: ...r to 7 14 Recording Operation Results of continuous measurement cannot be saved until the probe is removed During measurement measurement results cannot be saved to prevent unstable data from being st...

Страница 15: ...nt unit recording operation system setting about and factory reset Long press to enter the menu the operation is as follows Select from the menu 1 Up to select material 2 Down to select material 3 Con...

Страница 16: ...039in us to 0 394in us 1 Increase the value of sound velocity long press to speed up adjustment 2 Decrease the value of sound velocity long press to speed up adjustment 3 Save and return to menu 4 Re...

Страница 17: ...of each record 1 Page up 2 Page down 3 Enter the selected page 4 Return to the record interface 1 Up to choose functions 2 Down to choose functions 3 Enter the selected functions 4 Return to the menu...

Страница 18: ...cord details Enter the selected page press to select the record number and press to view the details details include record number current value maximum value minimum value and current value After vie...

Страница 19: ...the record to be deleted Deletion prompt In this interface the user can confirm whether to delete Use to select Yes No select Yes press to delete select No press to return to the interface to delete t...

Страница 20: ...n to select the function 3 Enter the selected function 4 Return to menu Select the Delete All function in the record operation interface the user can choose to delete all records Use to select Yes No...

Страница 21: ...light time In the system settings interface select the Shut function 7 18 shutdown time In the system settings interface select the Sound function the sound settings interface 1 Increase backlight tim...

Страница 22: ...setting 4 Return to system settings Select the Alarm function in the system setting interface to adjust the standard value first And press to adjust the tolerance limit Long press to increase the adju...

Страница 23: ...secondary measurement at the same point of the measured 8 1 Measurement method 1 select yes no 2 Select yes to restore factory settings select no to return to the menu 3 Return to menu Select in the...

Страница 24: ...be measured along the axis of the pipe or perpendicular to the axis of the pipe If the pipe diameter is long the measurement should be made in the direction of the vertical axis If the pipe diameter...

Страница 25: ...urface try to reduce the scratching of the probe on the surface During measurement at normal temperature the surface of the measured object should not exceed 60 C otherwise the probe cannot be used ag...

Страница 26: ...ormal phenomena such as the instrument is damaged and cannot be measured the liquid crystal display is abnormal the error is too large during normal use the keyboard operation fails or is confused etc...

Страница 27: ...glass Rubber custom 1 User define 1 m s User define 2 User define 3 custom 2 custom 3 sound velocity Aluminum common stainless Brass Copper Iron Cast iron Lead Nylon Silver Gold Zinc Brass Copper Iron...

Страница 28: ...adjust the angle between the ultrasonic probe s crosstalk spacer plate thin metal layer passing through the center of the bottom surface of the probe and the fine groove of the material to be measured...

Страница 29: ...ues of the probe crosstalk barrier plate perpendicular to the axis and parallel to the axis are read respectively the smaller of which is taken as the thickness measurement of the material at the meas...

Страница 30: ...and then it can be used as a test block For thin materials when its thickness is close to the lower limit of the probe s measurement a test block can be used to determine the exact lower limit Do not...

Страница 31: ...and then measure the workpiece at this speed of sound Casting measurement has its particularity The grain of the casting material is relatively coarse the structure is not dense enough and the measur...

Страница 32: ...to use the same material as the object to be tested and a test block with the same measurement direction as the object to be tested to calibrate the sound velocity of the material When measuring casti...

Страница 33: ...be replaced B 6 5 Multilayer materials composite materials It is not possible to measure multi layer materials with loosely bonded surfaces because ultrasonic waves cannot penetrate uncoupled surfaces...

Страница 34: ...operator should have the ability to identify abnormal readings usually rust spots corrosion pits and internal defects in the tested material will cause abnormal readings For solutions refer to the re...

Страница 35: ......

Страница 36: ...ices Hotline 400 125 6969 Tel 0769 85265688 Dongguan Sndway Electronic Co Ltd Add Sndway Science Technology Industrial Park 58 Tuanjie Road Humen 523930 Dongguan China Fax 0769 85116652 Url www sndway...

Отзывы: