Appendix A: Guidance and Manufacturer’s Declaration
A-4
Operation Manual SurgiVet
®
V3404 Plus
Recommended Separation Distances
The V3404 Plus is intended for use in an electromagnetic environment where radiated RF disturbances are
controlled. The customer or the user of the V3404 Plus can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the V3404 Plus as recommended below, according to the maximum output power of the communications
equipment.
The recommended separation distances between portable and mobile RF communication equipment and the
V3404 Plus is:
RATED MAXIMUM
OUTPUT POWER OF RF
TRANSMITTER (WATTS)
SEPARATION DISTANCE ACCORDING TO THE
FREQUENCY OF RF TRANSMITTER (METERS)
150 kHz to 80MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P
0.01
0.12
0.035
0.07
0.1
0.38
0.11
0.22
1
1.2
0.35
0.7
10
3.8
1.1
2.2
100
12
3.5
7.0
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where
P
is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
WARNING! The monitor should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the monitor should be observed to verify normal operation in the
configuration in which it will be used.
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......