Chapter 6: Oximetry
Operation Manual SurgiVet
®
V3404 Plus
6-5
Low Priority Alarm Messages
SpO2 SENSOR!
Indicates the SpO
2
sensor is disconnected from the monitor or the patient. Make sure the patient cable and
sensor connectors are all firmly seated. Make sure the sensor is positioned properly on the patient. If the
message persists, contact your authorized repair center.
SpO2 Signal
Indicates the oximeter circuitry could not measure the patient’s SpO
2
or pulse rate. Make sure the sensor is
positioned properly on the patient.
SpO2 Artifact
Indicates the presence of excessive artifacts. Check sensor positioning on the patient and restrict patient
movement.
Messages
SpO2 Search
Indicates the oximeter is adjusting to the patient’s signal. If this message persists, check for proper sensor
positioning on the patient.
Small Pulse
Indicates the oximeter signal strength is low. Try repositioning the sensor on the patient, or warming the
extremity to encourage blood flow.
Attaching the Patient - Oximetry
Choose the Sensor
PATIENT
SITE
CATALOG NO. & DESCRIPTION
Small/Medium Animal up
to 60 pounds
Pinna (ears), Toe
Webbing, Tongue
Rectum/Tail
Hock, Achilles
Tendon, etc.
V1711: Large ‘Y’ Sensor
V1702: Mini ‘Y’ Sensor
V3078: Mini Clip Sensor
V1700: Reflectance Sensor
V1710: Tail Wrap Sensor
V1707: Universal ‘C’ Sensor
Large Animals over 60
pounds
Pinna (ears), Toe
Webbing, Tongue
Rectum/Tail
Hock, Achilles
Tendon, etc.
V1711: Large ‘Y’ Sensor
V1702: Mini ‘Y’ Sensor
V3078: Mini Clip Sensor
V1700: Reflectance Sensor
V1710: Tail Wrap Sensor
V1707: Universal ‘C’ Sensor
Equine
Tongue
V1707L Universal ‘C’ Sensor
Содержание SurgiVet V3404 Plus
Страница 2: ......
Страница 8: ...Table of Contents vi Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 12: ...Warranty and Service Information Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 20: ...Chapter 1 Introduction 1 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 50: ...Chapter 6 Oximetry 6 10 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 64: ...Chapter 10 Serial Output 10 8 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 66: ...Chapter 11 Analog Output 11 2 Operation Manual SurgiVet V3404 Plus This page is intentionally left blank ...
Страница 84: ...Appendix B Revision History B 2 Autocorr Plus Operation Manual This page is intentionally left blank ...
Страница 85: ......